The Modulation Transfer Function (MTF) is a performance metric that measures the degradation of contrast in the image compared to contrast in the object. As contrast degrades, distinguishing small features in the image becomes increasingly harder for an observer. There are several factors that cause contrast degradation, such as diffraction effects, optical aberrations, and vignetting. Therefore, the MTF can be used to evaluate and compare optical systems.
To determine the system MTF, a sinusoidal pattern with perfect contrast is used as the object. Image contrast is defined as:
where Imax and Imin are maximum and minimum irradiances, respectively. The maximum value the contrast can take is 1, and the minimum value is 0. A cycle of the sinusoidal pattern corresponds to a black and a white fringe. Eash cycle is also referred to as a line pair (lp). Spatial frequency is often measured in line pairs or cycles per unit distance, which is typically mm. Thus, the spatial frequency unit is commonly represented as lp/mm or cycles/mm. For afocal systems, the units are converted to angular frequencies, typically measured in cycles/milliradian.
Figure 1: An optical system causes contrast degradation in the image with respect to contrast in the object.
Figure 1 shows the response of the system to a single frequency, however, the MTF measures the response for continuous frequencies up to the cutoff frequency fc. The cutoff frequency, which is the highest frequency the system can resolve, is given by:
The MTF is a function of a 2D spatial-frequency vector, but it is usually presented as a 1D plot. In this type of representation, it is useful to compare the response to that of a perfect system. In an aberration-free system, the contrast is only degraded by diffraction effects. The dotted line in Figure 2 shows the MTF for the aberration-free system, which is labeled as “Diffraction Limit”.
Figure 2: On-axis MTF plot of an optical system compared to the diffraction-limited MTF plot. The output image depicts excellent contrast at low frequencies and poor contrast at high frequencies, mimicking the MTF plot.
For off-axis field points, the MTF result depends on the orientation of the bar pattern. It is common to show the tangential (T) and radial directions (R) for each field point, as shown in Figure 3. In a tangential target, the bars are tangent to the image circle, and in a radial target, the bars are parallel to the radial direction.
Figure 3: Off-axis MTF plots for the tangential and radial directions. In CODE V, the target bars are relative to the Y axis. Radial MTF corresponds to vertical bars and tangential MTF corresponds to horizontal bars.
It is sometimes useful to measure through-focus MTF, which shows the MTF for selected frequencies as a function of defocus position. Another informative MTF chart is MTF vs field points, which shows the MTF for selected frequencies as a function of field points. Examples of these plots are shown in Figure 4.
Figure 4: Examples of MTF vs defocus position and MTF vs field charts.
So far, we have described sine-wave targets that represent a single frequency. The response of the system to a square-wave pattern is called Contrast Transfer Function (CTF). The CTF may be a more appropriate performance metric for systems that image objects with square-wave features, such as barcode readers. Computationally, the CTF can be approximated by summing a series of sine wave response values.
Figure 5: On-axis CTF. The target is a square-wave pattern instead of a sine-wave pattern.
The optical designer doesn’t necessarily aim to achieve the diffraction limit performance. The desired MTF curve is based on design requirements. The lens specifications are usually of the form of an MTF value for specific frequencies. The MTF specification may come, for instance, from the sensor pixel size and the Nyquist theorem, which describes the maximum spatial frequency that the sensor can resolve.
For example, assume a digital sensor with a pixel size of 7.4 μm x 7.4 μm. According to the Nyquist theorem, the highest frequency that can be resolved is 1 cycle/(2 x 7.4 um) = 0.0676 cycles/μm ≈ 68 cycles/mm. Based on the sensor characteristics, a typical performance requirement for the lens could be:
Figure 6 shows the MTF chart from a lens that would meet these requirements for all fields. It is important to know the sensor characteristics during the design process since in many cases the Nyquist frequency of the sensor is only a small fraction of the diffraction cutoff frequency fc.
Figure 6: System that meets requirements of MTF > 50% at 17 cycles/mm and MTF > 25% at 68 cycles/mm.
Another useful consideration is that for small aberrations, the degradation of the MTF occurs near half the diffraction cutoff frequency. This is because the MTF at zero frequency is always unity and the MTF at the diffraction cutoff frequency is zero. If we measure the aberrated MTF as a ratio of the aberration-free MTF, the result is a function that sags down in the middle.
Experimentally, there are different ways to measure the frequency response of an optical system. Two examples are the slanted-edge target and the three-bar target. These targets are easy to use since they just need to be imaged. The three-bar target is a common method to measure resolution, it consists of three parallel bars of certain width and separation that corresponds to a specific spatial frequency. To measure system resolution, multiple three-bar patterns at specific frequencies are arranged in a single chart, and the resulting image is inspected to determine the smallest features that can be resolved. A popular target with these characteristics is the USAF 1951 chart shown in Figure 7. The measured resolution from a three-bar target might differ from the predicted CTF, since the CTF assumes an infinitely long pattern of bars. In a pattern with a finite number of bars, there can be end effects that reduce the contrast of the bars at the ends of the pattern. For this reason, the measured contrast might be less than the predicted by the CTF.
Another method to experimentally measure the MTF is the slanted-edge target. This method is different in that frequency information can be obtained from a single target. The edge must be at a specified small angle with respect to the pixel array of the sensor, as shown in Figure 7. When the input is a step function instead of a point source, the irradiance distribution at the image is an edge spread function. The MTF can then be determined from the Fourier transform of the derivative of the edge spread function.
Figure 7: Examples of targets utilized to measure MTF experimentally.
To explain how the MTF is computed, first we need to define the Point Spread Function (PSF), the transfer function (TF), and their relationship with the MTF. The Point Spread Function (PSF) of an imaging system is the resulting irradiance distribution when the object is a point source. For coherent light, the PSF is related to the Fourier transform of the pupil function’s complex amplitude. For incoherent light, the PSF is related to the Fourier transform of the pupil function’s intensity. Therefore, the coherent and incoherent PSF relate as follows:
where is a 2D spatial-position vector.
The incoherent transfer function of a linear, shift-invariant system is given by the Fourier transform of the incoherent point spread function:
The Modulation Transfer Function (MTF) is the modulus of the optical transfer function:
Therefore, the MTF can be computed by calculating the complex autocorrelation of the pupil function, or by computing the Fourier transform of the incoherent PSF. Using the definition of the MTF, the OTF can be written in complex form as:
where is the phase. The MTF can only take positive values, but the OTF can be negative when there is phase reversal. For the image of the bar pattern, this would result in contrast reversal, meaning that white features become dark and dark features become light. In an MTF plot, the behavior is that the MTF curve reaches zero and then “bounces”. It is important to be aware of this behavior. After the bounce, the MTF technically increases even though it reached zero at a lower frequency. Another possible impact is that if the MTF is used in the error function during optimization, phase reversal may produce local minima.
Other common system specifications are Strehl ratio and RMS wavefront error. For small aberrations, specifying the Strehl ratio is equivalent to specifying the RMS wavefront error. The Strehl ratio is defined as the ratio of the peak intensity of the measured PSF to the peak intensity of the perfect PSF. The MTF is related to the inverse Fourier transform of the incoherent PSF. As a result, it is possible to express the Strehl ratio as the integral under the entire MTF curve, including frequencies beyond the Nyquist frequency (which are irrelevant). For this reason, in situations where Nyquist is well below the diffraction cutoff, it makes more sense to specify MTF at frequencies between zero and the Nyquist frequency, rather than specifying the Strehl ratio or RMS wavefront error.
CODE V is a powerful optical design tool. Its advanced analysis tools include the computation of the MTF by evaluating the autocorrelation of the pupil function, and the computation of the CTF by calculating a series summation of sine wave response values. In CODE V, MTF metrics are used throughout the design process for analyzing, optimizing, and tolerancing optical systems.